ACCS Practice Exam 2025 – Complete Test Preparation Guide

Question: 1 / 400

What action should be taken when there is a low pressure alarm on a ventilated patient with diminished breath sounds?

Apply supplemental oxygen

Use a bag-valve mask for ventilation

Use a tube exchanger to replace the ET tube

When approaching a low pressure alarm on a ventilated patient with diminished breath sounds, it is crucial to assess the potential causes of these alarms. A low pressure alarm often indicates a disconnection from the ventilatory circuit, a leak in the endotracheal tube cuff, or an inadequate seal, potentially leading to inadequate ventilation and oxygenation.

In this scenario, diminished breath sounds can suggest that there may be a significant airway issue, such as an incorrect placement or obstruction of the endotracheal tube. Using a tube exchanger to replace the ET tube is an appropriate action because it allows for re-establishing a secure airway quickly. This method minimizes the risk of extubation and can facilitate a more permanent solution if an issue exists in the current tube. It ensures the patient can be ventilated adequately and improves overall respiratory management.

While supplemental oxygen may provide short-term relief, it does not address the underlying cause of the low pressure and diminished breath sounds. Utilizing a bag-valve mask can aid in ventilation in an emergency but may not provide the control or protection needed in a critical situation. Increasing the respiratory rate without addressing the cause of the low pressure would not resolve the underlying problem and could lead to further complications. Therefore, replacing the

Get further explanation with Examzify DeepDiveBeta

Increase the respiratory rate on the ventilator

Next Question

Report this question

Subscribe

Get the latest from Examzify

You can unsubscribe at any time. Read our privacy policy